Language| CN | EN


Contact us

TEL :4008-266-224



Mobil:13510684211(Mr Luo)


Email:gc.luo@szlabpro.com

Your Current Location :Home » PRODUCT » Bridge Heating Circulators-BHC  » BHC 30
BHC 30

Name:BHC 30

Details
  • Appliance type [-] Bridge heating circulator
    Safety class according to DIN 12876 [-] III/FL
    Protection class according to DIN EN 60529 [-] IP20
    Working temperature range [°C] RT+10…220
    Temperature stability DIN 12876 [±K] 0.05
    Temperature display [-] LED
    Temperature control [-] Intelligent PID control
    Working temperature sensor [-] PT 100
    External temperature control [-] option
    Heating capacity [W] 3000
    Pump type [-] Pressure/ suction pump
    Pump capacity adjustable [-] no
    Pump pressure max. (0 liters discharge flow) [bar] 0.3
    Pressure pump (suction side) (0 liter flow) [bar] 0.2
    Pump pressure max. (0 bar back pressure) [L/min] 18
    Pump connection [-] M16x1
    Bath depth min. [mm] 150
    Dimension(WxDxH) [mm] (345…885) x 200 x 336
    Display resolution [°C] 0.1
    Warning function optical [-] yes
    Warning function acoustic [-] yes
    Warning function excess temperature [-] yes
    Warning function insufficient temperature [-] yes
    Sub-level protection [-] yes
    Warning function low filling volume [-] yes
    Safety temperature  range [°C] 0…260
    Calibration option [-] yes
    Device fixtion [-] Bridge
    Permissible ON time [%] 100
    Permissible ambient temperature [°C] 5..32
    Permissible relative humidity [%] 80
    RS 232 interface [-] option
    USB interface [-] option
    Weight [kg] 10
    Power supply [V] 200...230VAC/50Hz
    Power input [W] 3100
    Noise [db] ≤45
    Technical data complies with the standard [-] DIN 12876
  • Very easily mounted on different via variable bridge;
    powerful pressure/suction pump;
    Precise temperature control;
    Safety Class III (FL) for use with flammable fluids according to DIN 12876 standards;
    Convenient internal and external applications;
    Overtemperature and insufficient temperature warning funciton;
    Overtemperature  shut off avoid heating without liquid;
    Warning function low filling volume ;
    1-3 points calibration function;
    Warning function optical  and acoustic;
    Adjustable safety temperature;
    Option interface RS232 and USB;

Links