Language| CN | EN


Contact us

TEL :4008-266-224



Mobil:13510684211(Mr Luo)


Email:gc.luo@szlabpro.com

Your Current Location :Home » PRODUCT » Combined Circulators-CC  » CC-0520E
CC-0520E

Name:CC-0520E

Details
  • Working temperature range

    [°C]

    -5..100

    Temperature stability DIN 12876

    [±K]

    ±0.050.1

    Temperature display

    [-]

    LED

    Working temperature sensor

    [-]

    PT 100

    External temperature control

    [-]

    no

    Heating capacity

    [W]

    2000

    Pump type

    [-]

    Pressure/ suction pump

    Pump capacity adjustable

    [-]

    no

    Pump pressure max.

    [bar]

    0.3

    Pressure suction max.

    [bar]

    0.2

    Pump pressure max.

    [L/min]

    22

    Pump connection

    [-]

    M16x1

    Bath  volume

    [L]

    20.0

    Bath opening(WxDxH)
    [mm]
         295X297X150

    Dimension(WxDxH)

    [mm]

    350X570X778

    Display resolution

    [°C]

    0.1

    Device fixtion

    [-]

    Stand floor

    Permissible ON time

    [%]

    100

    Permissible ambient temperature

    [°C]

    5..32

    Permissible relative humidity

    [%]

    80

    Power supply

    [V]

    200...230VAC/50Hz


  • Key features

    1.   Splash-proof keypad, Large, bright LED display

    2.   Compact construction, easily installed on the laboratory table and fume hood

    3.   The removable venting grid for an easy cleaning of the condenser

    4.   6 segments program control (Optional).

    5.   Powerful immersion pressure/suction pump, suitable for continuous operation and for open and closed external application

    6.   Liquid level protection, optical and acoustic safety warning.

    7.   High-performance and high-quality parts, the units with stable quality and reliable performance.

    8.   Stainless steel water bath with  big opening

    9.   Natural refrigerant

    10.  1-2 points temperature calibration function;

    11. Optional external temperature sensor for  external applications.

          12.  Optional Connection of a magnetic valve possible

          13.  Optional RS485/RS232 interface  for external remote control


Links